Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0628-2011

ASP Automatic Endoscope Reprocessor (AER)

Advanced Sterilization Products / initiated 2010-02-19 / Terminated / root cause: sterility

Advanced Sterilization Products began a recall of ASP Automatic Endoscope Reprocessor (AER), Product Code: 20300 & 20301 The AER is a machine designed to automatically wash and high-level disinfect flexible... on . The reason FDA records is "The recall was initiated because Advanced Sterilization Products (ASP) received three customer reports of elevated temperature of the high-level disinfectant solution used in the ASP AER as a result of a possible...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0628-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Sterilization Products
Product
ASP Automatic Endoscope Reprocessor (AER), Product Code: 20300 & 20301 The AER is a machine designed to automatically wash and high-level disinfect flexible, submersible fiberoptic or video endoscopes.
Product code
KOG
Reason for recall
The recall was initiated because Advanced Sterilization Products (ASP) received three customer reports of elevated temperature of the high-level disinfectant solution used in the ASP AER as a result of a possible temperature monitoring system malfunction. This recall communication advised consignees that if high-level disinfectants are exposed to elevated temperatures, fumes may occur and resul
Root cause tag
sterility
FDA root cause
Pending
Action
Advanced Sterilization Products sent an URGENT: PRODUCT CORRECTION letter dated February 19, 2010, to all customers. The letter identified the product, the problem, and the action to be taken by the customer. The recall communication was initiated with ASP forwarding a "Urgent: Product Correction" letter with attached customer fax-back response form to the affected customers. Customers were also instructed to complete and sign the enclosed customer fax-back response form indicating that they have disconnected the heater on the ASP AER according to the instructions provided and fax the form to 1-888-345-1435. If customers have any questions or need assistance, they were instructed to call ASP Professional Services at 888-783-7723, option 3 then option 1.
Quantity in commerce
1,881 units (1514 in US)
Distribution
Worldwide Distribution -- United States, Argentina Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, Hungary, Mexico, Panama, Guatemala, Honduras, Poland, Puerto Rico, Virgin Islands, Singapore, Taiwan, Venezuela, Croatia, and Bulgaria.
Product codes and lots
Product Code: 20300 & 20301