Z-0445-2011

Lumen Biomedical Xtract Aspiration Catheter system, intravascular catheter, consisting of One Catheter, Two...

Volcano Corporation / initiated 2010-11-03 / Terminated / root cause: design

Volcano Corporation began a recall of Lumen Biomedical Xtract Aspiration Catheter system, intravascular catheter, consisting of One Catheter, Two 30 mL Syringes, One Extension Tube with Stopcock... on . The reason FDA records is "Potential for product failure. Packaging/Storage issue results in compromised integrity and potential product failure during use, leading to potential patient injury". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0445-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Volcano Corporation
Product
Lumen Biomedical Xtract Aspiration Catheter system, intravascular catheter, consisting of One Catheter, Two 30 mL Syringes, One Extension Tube with Stopcock, and One 40 um Cell Strainer. Single-user, 0.014" guidewire compatible intravascular extraction and aspiration catheter, Part number 1450-0001 size 4.2F, and part number 1451-003 size 5.8F, distributed by Volcano Corporation.
Product code
OGX
Reason for recall
Potential for product failure. Packaging/Storage issue results in compromised integrity and potential product failure during use, leading to potential patient injury.
Root cause tag
design
FDA root cause
Device Design
Action
Firm initiated recall on 11/03/2010, and sent notifications to its consignees via letter. The firm has requested that consignees provide the notification to all personnel within the organization and letters will be sent directly to known customers on 11/04/2010 by Volcano. Customers have been instructed to cease use and contact customer service to arrange returns, and credit will be issued.
Quantity in commerce
4724 units
Distribution
Worldwide distribution: USA, Greece, Egypt, Kuwait, Netherlands, Saudi Arabia, Italy, France, Sweden, Portugal, Great Britain, Austria, Switzerland, Germany, Romania, Poland, and Bulgaria
Product codes and lots
Part number 1450-001 size 4.2F and 1451-003 size 5.8F, all lots.