Z-0719-2011

Bard CritiCore Monitor, Software Version 2.1, Catalog number 000002N, C.R. Bard, Inc., Covington, GA 30014...

C.R. Bard Inc / initiated 2010-11-17 / Terminated

C.R. Bard, Inc., Urological Division began a recall of Bard CritiCore Monitor, Software Version 2.1, Catalog number 000002N, C.R. Bard, Inc., Covington, GA 30014. Intended for multiple uses as it is an electronic... on . The reason FDA records is "Error can occur in the reading of the Present Interval Output (PIO) reading when the device is subjected to bumping or forceful replacement of the fluid container". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0719-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.R. Bard Inc
Product
Bard CritiCore Monitor, Software Version 2.1, Catalog number 000002N, C.R. Bard, Inc., Covington, GA 30014. Intended for multiple uses as it is an electronic battery-operated monitor used in conjunction with a single-use disposable urine collection container and/or a temperature sensing Foley catheter
Product code
EXS Urinometer, Electrical
Reason for recall
Error can occur in the reading of the Present Interval Output (PIO) reading when the device is subjected to bumping or forceful replacement of the fluid container.
FDA root cause
Other
Action
Consignees were notified by an Urgent: Product Advisory Letter and Medica Device Recall letter on/about November 18, 2010. The letter identified the affected product, stated the reason for recall, the interim precautions, and actions to be taken. The letter requested acknowledgement of the recall and to return the entire affected product. Each consignee that is a distributor was instructed to forward the letter to their consignees to retrieve relevant product from those locations. Affected product received by Bard Medical will be reworked with the software upgrade and place back into distribution. Consignees will either receive a software upgrade on site by a third party or replacement product with the software upgrade. For international distribution, the hospital will be informed of this issue through Bard affiliate and Notified Body Bard Europe and managed according to the requirement. If there are questions or assistance is needed, customers can contact BD at 1-770-784-6969.
Quantity in commerce
234 units
Distribution
Worldwide Distribution -- U.S., Netherlands, and Japan.
Product codes and lots
Catalog number 000002N, Serial numbers: DYTA0001, DYTA0002, DYTA0003, DYTA0007, DYTA0008, DYTA0011, DYTA0012, DYTA0015, DYTA0016, DYTA0017, DYTA0018, DYTA0019, DYTA0020, DYTA0021, DYTA0022, DYTA0024, DYTA0025, DYTA0027, DYTA0028, DYTA0029, DYTA0030, DYTA0031, DYTA0032, DYTA0033, DYTA0034, DYTA0036, DYTA0037, DYTA0038, DYTA0039, DYTA0040, DYTA0041, DYTA0042, DYTA0043, DYTA0044, DYTA0045, DYTA0046, DYTA0047, DYTA0049, DYTA0050, DYTA0051, DYTA0052, DYTA0053, DYTA0054, DYTA0056, DYTA0057, DYTA0058, DYTA0059, DYTA0060, DYTB0001, DYTB0003, DYTB0004, DYTB0005, DYTB0009, DYTB0010, DYTB0011, DYTB0012, DYTB0013, DYTB0015, DYTB0016, DYTB0020, DYTB0021, DYTB0022, DYTB0023, DYTB0024, DYTB0025, DYTB0026, DYTB0027, DYTB0029, DYTB0030, DYTB0031, DYTB0033, DYTB0038, DYTC0001, DYTC0004, DYTC0010, DYTC0011, DYTC0016, DYTC0017, DYTC0018, DYTC0019, DYTC0020, DYTC0021, DYTC0022, DYTC0024, DYTC0025, DYTC0026, DYTC0027, DYTC0029, DYTC0030, DYTC0031, DYTC0032, DYTC0033, DYTC0034, DYTC0035, DYTC0036, DYTC0037, DYTC0038, DYTC0039, DYTC0040, DYUC0001, DYUC0002, DYUC0004, DYUC0005, DYUC0009, DYUC0012, DYUC0014, DYUC0015, DYUC0016, DYUC0017, DYUC0018, DYUC0019, DYUC0020, DYUD0001, DYUD0002, DYUD0003, DYUD0004, DYUD0005, DYUD0006, DYUD0007, DYUD0008, DYUD0009, DYUD0011, DYUD0012, DYUD0013, DYUD0014, DYUD0015, DYUD0016, DYUD0017, DYUD0019, DYUD0020, DYUD0021, DYUD0022, DYUD0023, DYUD0026, DYUD0027, DYUD0028, DYUD0029, DYUD0030, DYUD0031, DYUD0033, DYUD0034, DYUD0035, DYUD0036, DYUD0037, DYUD0038, DYUD0039, DYUD0042, DYUD0043, DYUD0045, DYUD0046, DYUD0047, DYUD0048, DYUD0049, DYUD0050, DYUD0052, DYUD0053, DYUD0055, DYUD0057, DYUD0058, DYUD0059, DYUE0002, DYUE0007, DYUE0008, DYUE0011, and DYUE0014.