Z-0815-2011

CuttingEdge Advantage Rasp Handle, Straight Rasp Handle, & Specialty Head Center Accolade Broach Handle...

Howmedica Osteonics Corp / initiated 2010-11-01 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of CuttingEdge Advantage Rasp Handle, Straight Rasp Handle, & Specialty Head Center Accolade Broach Handle; Stryker Howmedica Osteonics, Non-sterile Instruments... on . The reason FDA records is "The CuttingEdge Advantage Rasp Handle, Straight Rasp Handle, and Specialty Head Center Accolade Broach Handle may fracture in two pieces along the width of the body under the striking plate. This may extend surgical...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0815-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
CuttingEdge Advantage Rasp Handle, Straight Rasp Handle, & Specialty Head Center Accolade Broach Handle; Stryker Howmedica Osteonics, Non-sterile Instruments; Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430;
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The CuttingEdge Advantage Rasp Handle, Straight Rasp Handle, and Specialty Head Center Accolade Broach Handle may fracture in two pieces along the width of the body under the striking plate. This may extend surgical time under anesthesia, cause trauma to the surgeon or patient, or contaminate the surgical site.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker issued a Product Correction Bulletin dated November 2010 to their consignees, alerting surgeons of the need to inspect the instruments before and after each surgical use, and, if a defect is detected, to contact Stryker for return of the product. Stryker can be contacted at 201 831-5288 concerning this issue. In response to their revised strategy, Stryker issued notification letters and product recall acknowledgement forms to their branches and agencies on 2/17/2011 for 2/18/2011 delivery via Fed Ex and to their hospital risk managers on 2/18/2011 for 2/21/2011 delivery via Fed ex.
Quantity in commerce
459 units
Distribution
Worldwide Distribution: Throughout the US, the countries of Canada and Japan, and the regions of Asia Pacific, Europe, Middle East, Africa, and Latin America.
Product codes and lots
CuttingEdge Advantage Rasp Handle - Catalog number 1100-1000; All lot codes  Straight Rasp Handle - Catalog number 1440-1400; All lot codes  Specialty Head Center Accolade Broach Handle - Catalog number I-H088HF00, I-H1389HF00 ; All lot codes