Z-0844-2011

Fresenius Medical CombiSet BVM Hemodialysis Blood Tubing Set P/N: 03-2795-7

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2010-11-19 / Terminated / root cause: design

Fresenius Medical Care Holdings, Inc. began a recall of Fresenius Medical CombiSet BVM Hemodialysis Blood Tubing Set P/N: 03-2795-7 on . The reason FDA records is "Hemodialysis Blood tubing Sets may develop kinking of the arterial line". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0844-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius Medical CombiSet BVM Hemodialysis Blood Tubing Set P/N: 03-2795-7
Product code
FJK Set, Tubing, Blood, With And Without Anti-Regurgitation Valve
Reason for recall
Hemodialysis Blood tubing Sets may develop kinking of the arterial line.
Root cause tag
design
FDA root cause
Device Design
Action
Fresenius Medical issued recall notification dated 11/19/10, "URGENT RECALL FMCNA Combiset BVM Hemodialysis Blood Tubing Sets (Part Numbers 03-2695-9 and 03-27957)", by certified mail with return receipt. Customers have been instructed to check their stock immediately to determine if they have any affected lots on hand. If customers have the affected lots, they have been instructed to discontinue use immediately and place the product in a secure segregated area for return to Fresenius Medical Care North America. Customers have been instructed to contact Fresenius Medical Care Customer Service Team Representative for instructions on how to return the recalled product.
Quantity in commerce
10, 032 units (418 cases x 24 units)
Distribution
Distribution Nationwide and Canada
Product codes and lots
Lots 09JR01174, 09JR01229, 09NR01139, 10KR01801