Z-1333-2011

Winged Infusion Set with 90 degree Huber Needle and Y Injection Site Distributed By: FloridaInfustion...

Vygon Mfg / initiated 2010-11-16 / Terminated / root cause: supplier

Churchill Medical Systems, Inc. began a recall of Winged Infusion Set with 90 degree Huber Needle and Y Injection Site Distributed By: FloridaInfustion Services, Inc., Palm Harbor, FL 34683 Item # FIY-2210... on . The reason FDA records is "Routine testing conducted to challenge the success of sterilization was not fully completed for sterile products". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1333-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Vygon Mfg
Product
Winged Infusion Set with 90 degree Huber Needle and Y Injection Site Distributed By: FloridaInfustion Services, Inc., Palm Harbor, FL 34683 Item # FIY-2210 Intended use: Intravascular Administration Set
Product code
FPA Set, Administration, Intravascular
Reason for recall
Routine testing conducted to challenge the success of sterilization was not fully completed for sterile products.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
The firm, Churchill Medical/Vygon, sent an "URGENT MEDICAL DEVICE RECALL" letter dated November 16, 2010, via overnight mail to all customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed to check their stock for affected product; immediately cease use and distribution and quaratine the affected product; contact Vygon's Customer Service Department at 1-800-473-5414 for a return Merchandise Authorization number; complete and return the Recall Acknowledgement and Inventory Return Form via fax to 215-390-2019 and mail return product with form to: Vygon, 103A Park Drive, Montgomeryville, PA 18936, and if the customer is a distributor, they were to provide their customers with a copy of this notification and attached form. Note: An additional follow up communication will be provided to the customers after reconciliation of the returned inventory and all inventory will be credited or replaced. If you have any questions, please contact your local sales representative or Vygon's Customer Service Department at 1-800-473-5414, or by email at [email protected].
Quantity in commerce
1380 units
Distribution
Nationwide Distribution: USA including states of: CO, IA, OH, MA, MD, MO, NY, FL, SC, FL, and VA.
Product codes and lots
Lot Number: 1010034