Accessory Cart for The ACHIEVA 3.0T and Panorama 1T Magnetic Resonance Imaging unit. Consists of diagnostic...
Philips Medical Systems North America Co. Phillips began a recall of Accessory Cart for The ACHIEVA 3.0T and Panorama 1T Magnetic Resonance Imaging unit. Consists of diagnostic devices that produce cross sectional images... on . The reason FDA records is "The Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0794-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Accessory Cart for The ACHIEVA 3.0T and Panorama 1T Magnetic Resonance Imaging unit. Consists of diagnostic devices that produce cross sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- The Accessory Cart contained magnetic material and is attracted to the magnet when bringing it close to the system.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 05/12/2009, Philips began sending out the URGENT-Field Safety Notice MR Accessory Cart to their consignees. The letter identified the affected product and stated that the drawer rails were made from magnetic material and were not according specification. The hazards involved were also provided along with how to identify the affected product. The firm issued the FCO 78100296 and the firm's Field Service Engineer replaced the magnetic rails with the non-magnetic rails at the consignee's site. The consignees can contact contact the Philips Healthcare Call Center at 800-722-9377 #5 (then follow prompts) and reference FCO 78100296 with any communication.
- Quantity in commerce
- 3 units
- Distribution
- Nationwide Distribution -- IL, MI, and MD.
- Product codes and lots
- Site numbers: 505048, 505705, and 41445594.