Z-1739-2011

Dimension Vista(R) Gentamicin (GENT) Flex(R) reagent cartridge, K4012 This method is an in vitro diagnostic...

Siemens Healthcare Diagnostics Inc. / initiated 2010-11-30 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista(R) Gentamicin (GENT) Flex(R) reagent cartridge, K4012 This method is an in vitro diagnostic test for the quantitative measurement of... on . The reason FDA records is "False elevation of Dimension Vista(R) Gentamicin results for patients and QC when prior reagent aspirated by the reagent probe is any of the following plasma protein methods: ASL, B2MIC, B2MU, CRP,CYSC, HCYS, hcCRP...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1739-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista(R) Gentamicin (GENT) Flex(R) reagent cartridge, K4012 This method is an in vitro diagnostic test for the quantitative measurement of gentamicin, an aminoglycoside antibiotic, in human serum and plasma on the Dimension Vista System. Gentamicin measurements may be used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.
Product code
LCD Enzyme Immunoassay, Gentamicin
Reason for recall
False elevation of Dimension Vista(R) Gentamicin results for patients and QC when prior reagent aspirated by the reagent probe is any of the following plasma protein methods: ASL, B2MIC, B2MU, CRP,CYSC, HCYS, hcCRP, IGE, IGG3, IGG4, RF, and STFR resulting in inappropriate patient treatment.
FDA root cause
Other
Action
Siemens Healthcare Diagnostics, Inc. sent an Urgent Field Safety Notice letter dated November, 2010, to all affected customers. Customers were requested to run GENT on a Dimension Vista(R) system that does not run the listed plasma protein methods. Instructions were provided for disabling methods on the instrument to prevent inadvertently testing Gent along with plasma protein methods. Customers were instructed to : Forward the notification to anyone to whom they may have distributed the product. Complete and return the Field Correction Effectiveness Check Form to Siemens Healthcare Diagnostics at (302) 631-8467. For further information contact Siemens Technical Solutions Center at 1-800-441-9250 or their local Siemens representative.
Quantity in commerce
N/A
Distribution
Nationwide Distribution including AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Product codes and lots
All Lots