Boston Scientific brand Matrix2 Helical Soft SR Coils. Intended for embolization of certain intracranial...
Boston Scientific Corporation, Neurovascular, 47900 Bayside Parkway, Fremont, CA 94538 began a recall of Boston Scientific brand Matrix2 Helical Soft SR Coils. Intended for embolization of certain intracranial aneurysms and for arterial and venous embolizations in... on . The reason FDA records is "The product appears to have premature degradation of the PGLA coating on the Matrix Coil". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0855-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific brand Matrix2 Helical Soft SR Coils. Intended for embolization of certain intracranial aneurysms and for arterial and venous embolizations in the peripheral vasculature.
- Product code
- HGC Clamp, Uterine
- Reason for recall
- The product appears to have premature degradation of the PGLA coating on the Matrix Coil.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- US consignees were notified by letter sent by Federal Express on October 19, 2010. Customers out of the United States were notified by Federal Express if possible and by personal delivery by locally assigned sales representatives as needed.
- Quantity in commerce
- 77,364 total units
- Distribution
- Worldwide distribution. USA, Argentina, Austria, Australia, Belgium, Bahrain, Brazil, Canada, Switzerland, Chile, China, Colombia, Serbia, Czech Republic, Germany, Denmark, Spain, Finland, France, Great Britain, Greece, Hong Kong, Hungary, Ireland, India, Israel, Italy, Jordan, Japan, South Korea, Lithuania, Latvia, Macau, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovakia, Thailand, Taiwan, Turkey, Ukraine, Vietnam, and Yemen.
- Product codes and lots
- Material Product Description Catalog # Batch Numbers Expiry Date M003471204SRO Matrix2¿ Helical Soft SR Coils 471204-SR 11297523,12469527 Oct2010 through 2mm x4cm Feb2012 M003471206SRO Matrix2¿ Helical Soft SR Coils 471206-SR 11422256,11697451,11952856,12361059 Dec2010 through 2mm x6cm Dec2011 M003471208SRO Matrix2¿ Helical Soft SR Coils 471208-SR 11389025,11552556,12514612,13606645 Dec2010 through 2mm x8cm Jun2013
FIRST ARTICLE / WEEKLY