Z-1971-2011

Precision 500D X-ray systems with under-table collimator part number 5234960. The Precision 500D is designed...

GE Healthcare / initiated 2010-06-10 / Terminated

GE Healthcare, LLC began a recall of Precision 500D X-ray systems with under-table collimator part number 5234960. The Precision 500D is designed to perform radiographic and fluoroscopic x-ray... on . The reason FDA records is "Use of high force when moving the image carriage/IDD into forward/park position can cause changes to the mechanical alignment of the collimator blades and allow the blades to open wider than specified, consequently...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1971-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Precision 500D X-ray systems with under-table collimator part number 5234960. The Precision 500D is designed to perform radiographic and fluoroscopic x-ray examinations.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
Use of high force when moving the image carriage/IDD into forward/park position can cause changes to the mechanical alignment of the collimator blades and allow the blades to open wider than specified, consequently allowing radiation outside the permitted field without warning to the user. This issue manifests with symptoms wherein the user cannot fully close down the lateral blades.
FDA root cause
Radiation Control for Health and Safety Act
Action
GE Healthcare sent an Importan Electronic Product Radiation Warning letter to customers. The letter discussed the product defect description and related hazards, affected product details, instructions, and product correction. The firm will send a GE Field Service Representative to visit each affected customer site to inspect and correct all potentially affected units in the field following the instructions in field modification instructions No: 10846. If you have any questions please call 800-437-1171.
Quantity in commerce
440 US
Distribution
Nationwide Distribution
Product codes and lots
Part number 5234960.