Z-1093-2011

Hemochron Whole Blood Coagulation Systems Activated Clotting Time (ACT); Package label states -International...

Accriva Diagnostics, Inc. / initiated 2010-11-30 / Terminated / root cause: manufacturing process

International Technidyne Corp. began a recall of Hemochron Whole Blood Coagulation Systems Activated Clotting Time (ACT); Package label states -International Technidyne Corporation Response Celite ACT Tube... on . The reason FDA records is "Some Activated Clotting Time (ACT) Tubes may contain a component that does not meet its specification, making it possible for a failure to detect a tube or the presence of a clot". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1093-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Accriva Diagnostics, Inc.
Product
Hemochron Whole Blood Coagulation Systems Activated Clotting Time (ACT); Package label states -International Technidyne Corporation Response Celite ACT Tube HRFTCA510; Lot number and expiration date are listed. Each box contains 95 ACT tubes tests. Used for heparin anticoagulation monitoring during cardiopulmonary bypass surgery, percutaneous transluminal coronary angioplasty (PTCA), interventional radiology, extracorporeal membrane oxygenation (ECMO), hemofiltration, hemodialysis and critical care.
Product code
KQG Instrument, Coagulation
Reason for recall
Some Activated Clotting Time (ACT) Tubes may contain a component that does not meet its specification, making it possible for a failure to detect a tube or the presence of a clot.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
ITC notified their distributors and end user customers with an Urgent Medical Device Recall letter dated November 29, 2010 through UPS overnight mail. The letter identified the affected product and explained the reason for the recall. Distributors and customers should check their inventory for any affected product. Distributors should immediately discontinue shipment of the affected product and place the product on hold. Then regardless of whether they have any remaining inventory, they should complete the enclosed Distributor Account Tracking Form. The firm will contact them to arrange for any affected product to be returned and will credit them for any unused, returned product. Customers are to stop using the affected product and remove it from their inventory. They should also complete the attached Customer Account Tracking Form and return it to the firm. The firm will contact them to arrange for the return of any affected inventory and will credit the customers for any unused, returned product. If customers believe that they have experienced a problem with the product, they should take any additional actions regarding their records or patients. Questions should be directed to ITC Nexus Dx Technical Support at 800-631-5945 (US) or 732-548-5700 (International) or e-mail [email protected].
Quantity in commerce
265 boxes (95 ACT Tubes per box)
Distribution
Worldwide Distribution
Product codes and lots
Lot Number G0FTE157A, Expiration Date 7/31/2013 Lot Number G0FTE158A, Expiration Date 7/31/2013 Lot Number G0FTE159A, Expiration Date 7/31/2013 Lot Number G0FTE159B, Expiration Date 7/31/2013