Z-1474-2011

Aesculap MicroSpeed Uni Control Box, Part #GD670 Intended for high speed cutting, sawing, drilling and...

Aesculap Inc / initiated 2010-11-22 / Terminated / root cause: software

Aesculap Inc began a recall of Aesculap MicroSpeed Uni Control Box, Part #GD670 Intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of... on . The reason FDA records is "Potential safety issue in that the control box can misinterpret the information sent from the low speed motor if the control magnet inside the motor changes its posthion". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1474-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
Aesculap MicroSpeed Uni Control Box, Part #GD670 Intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro and Maxillo facial Surgery.
Product code
HBC Motor, Drill, Electric
Reason for recall
Potential safety issue in that the control box can misinterpret the information sent from the low speed motor if the control magnet inside the motor changes its posthion.
Root cause tag
software
FDA root cause
Software design
Action
AESCULAP sent an IMPORTANT CORRECTION NOTICE letter dated November 22, 2010, to all affected consignees. The letter identified the product, the problem, and the action the consignee should take. Consignees were instructed to not use the motors for thread cutting and removal of screws. If any unusual reactions of any cutting tools were observed consignees should check to ensure that the motor is not running in the opposite direction. If any error message is displayed, consigness were instructed to not use the motor. Consignees were also instructed to check both directions of running before use and do not use a motor which produces an error message in one direction. Aesculap would be providing all customers with a software update to the MicroSpeed Uni control box either by the unit being returned to Aesculap for update or by firm representative doing the upgrade on site at their facility. For questions regarding this recall call 800-258-1946 x5254 or 800-258-1946 x5267.
Quantity in commerce
91 units
Distribution
Worldwide Distribution - USA including CA, CO, DE, FL, GA, IN, LA, MO, NC, NJ, OK, PA, TN, TX, and VA and the country of Canada.
Product codes and lots
Part #GD670 - all distributed lot numbers and serial numbers