Z-1477-2011

ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows

Siemens Healthcare Diagnostics Inc. / initiated 2010-12-14 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc began a recall of ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows: Product code: 09427226 (128532) 100 test kit Product code: 01678114 (128533) 500... on . The reason FDA records is "Positive bias (40-55%) with patient results greater than the upper limit of normal and less than the upper limit of normal". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1477-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Centaur CA 125 II Assay packaged in 100 and 500 test kits as follows: Product code: 09427226 (128532) 100 test kit Product code: 01678114 (128533) 500 Test kit
Product code
LTK Test, Epithelial Ovarian Tumor-Associated Antigen (Ca125)
Reason for recall
Positive bias (40-55%) with patient results greater than the upper limit of normal and less than the upper limit of normal
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens Healthcare Diagnostics on 12 /14/10 issued Urgent Device Recall notice via Federal Express to accounts. This notice informs the customer of the recall and the actions that need to be taken. The Urgent Device Recall notice directs customers to immediately discontinue the use of ADVIA Centaur CA 125 II assay kits that contain CA 125 II reagent lot 144. It also recommends that patient results obtained with this reagent lot be reviewed. A confirmation fax-back form is included to ensure that all customers worldwide have been notified and have complied with the mandatory action. Technical questions, please contact the Technical Solutions Center at 1-877-229-3711
Quantity in commerce
2960 units
Distribution
Worldwide distribution, including USA, Canada, Argentina, Austria, Belgium, China, Denmark, France, Germany, Great Britain, Greece, India, Italy, Japan, Korea, Lavitia, Lithuania, Malyasia, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Romania, Saudi Arabia, Switzerland, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Czech Republic, Taiwan, Turkey, Uganda, and Venezuela
Product codes and lots
Lot Code: 144 Affected kit lot numbers: 62431144, 63124144, 63519144, 63876144, 62450144 and 63822144 Expiry Date: 15 September 2011