Z-1398-2011

Abbott RealTime HIV-1 Assay, List 6L18, in vitro diagnostic; Abbott Molecular Inc., Des Plaines, IL 60018 USA...

Abbott Molecular, Inc. / initiated 2010-12-20 / Terminated / root cause: labeling and UDI

Abbott Molecular began a recall of Abbott RealTime HIV-1 Assay, List 6L18, in vitro diagnostic; Abbott Molecular Inc., Des Plaines, IL 60018 USA The Abbott RealTime HIV-1 assay is an in vitro... on . The reason FDA records is "Internal testing identified the potential of under filled and over filled enzyme reagent vials which may result in obtaining less than the total number of tests per kit as described on the package insert". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1398-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Molecular, Inc.
Product
Abbott RealTime HIV-1 Assay, List 6L18, in vitro diagnostic; Abbott Molecular Inc., Des Plaines, IL 60018 USA The Abbott RealTime HIV-1 assay is an in vitro reverse transcription-polymerase chain reaction (RT-PCR) assay for the quantitation of Human Immunodeficiency Virus type 1 (HIV-1) on the automated m2000 System in human plasma from HIV-1 infected individuals over the range of 40 to 10,000,000 copies/mL. The Abbott RealTime HIV-1 assay is intended for use in conjunction with clinical presentation and other laboratory markers for disease prognosis and for use as an aid in assessing viral response to antiretroviral treatment as measured by changes in plasma HIV-1 RNA levels. This assay is not intended to be used as a donor screening test for HIV-1 or as a diagnostic test to confirm the presence of HIV-1 infection.
Product code
MTL Monitor,Test,Hiv-1
Reason for recall
Internal testing identified the potential of under filled and over filled enzyme reagent vials which may result in obtaining less than the total number of tests per kit as described on the package insert.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Abbott Molecular sent a Field Advisory Notice letter to all RealTime HIV-1, CT/NG and HBV assay customers on December 20, 2010, informing them that internal testing identified the potential of under filled and over filled enzyme vials which may result in obtaining less than the total number of tests per kit as described in the package insert. The customers were told that specimens that meet acceptance criteria, per the product labeling are not affected by the situation and are considered valid results, and that the issue has no impact on results for those runs that generate valid results. Any questions were directed to Abbott Technical Support at 1-800-553-7042, Option 2.
Distribution
Nationwide Distribution
Product codes and lots
All in date lots