PolyAx Locking Plate System, POLYAX FEMORAL PLATE - LEFT, 12 hole, Sterile, REF 8141-31-112, Depuy...
Depuy Orthopaedics, Inc. began a recall of PolyAx Locking Plate System, POLYAX FEMORAL PLATE - LEFT, 12 hole, Sterile, REF 8141-31-112, Depuy Orthopaedics, Inc., Warsaw, IN on . The reason FDA records is "The firm determined that the 8.0 mm hole threads in the PolyAx Distal Femoral Locking Plate were incorrectly manufactured, which prevents the mating screw from locking in place. The screw should lock with the plate...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1584-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- PolyAx Locking Plate System, POLYAX FEMORAL PLATE - LEFT, 12 hole, Sterile, REF 8141-31-112, Depuy Orthopaedics, Inc., Warsaw, IN.
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- The firm determined that the 8.0 mm hole threads in the PolyAx Distal Femoral Locking Plate were incorrectly manufactured, which prevents the mating screw from locking in place. The screw should lock with the plate, instead the screw passes through the plate.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Urgent Information - Recall Notice letters, dated November 1, 2010 were sent to customers. The letters described the reason for recall and identified the product affected along with the affected units. The letters also provided clinical guidance and stated that the problem had been corrected and that DePuy was in the process of replenishing its inventory. Hospitals were to determine if any of the recalled components they received were still on hand, and return them immediately to their DePuy Representative or directly to DePuy for credit.
- Quantity in commerce
- 246 all products
- Distribution
- Nationwide Distribution -- ME, MA, FL, VA, CO, TX, WY, AZ, MO, WI, GA, PA, AL, NY, OH, NJ, NC, and CA.
- Product codes and lots
- DLHBGZ, DLHCRD, DLIB44.