BD Visitec High Viscosity Injector 4mm .90 (20G) [Hammer] REF 585173 High Viscosity Injectors are used during...
Becton Dickinson & Co / initiated 2010-12-17 / Terminated
Becton Dickinson and Company began a recall of BD Visitec High Viscosity Injector 4mm .90 (20G) [Hammer] REF 585173 High Viscosity Injectors are used during retinal surgery for fluid infusion. A High... on . The reason FDA records is "Restricted or no fluid flow for Visitec High Viscosity Injector". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-2159-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Visitec High Viscosity Injector 4mm .90 (20G) [Hammer] REF 585173 High Viscosity Injectors are used during retinal surgery for fluid infusion. A High Viscosity Injector is placed into the sclera and temporarily sutured into place.
- Product code
- HMX Cannula, Ophthalmic
- Reason for recall
- Restricted or no fluid flow for Visitec High Viscosity Injector.
- FDA root cause
- Component change control
- Action
- Beaver Visitec (BVI) notified customers via telephone and by sending an Urgent Recall Customer Notification letter dated December 7, 2010. The letter identified the product, the problem, and the action to be taken by the customer. , Customers were requested to return all unused product in their inventory for credit or refund and complete the attached Customer Response Form and return it by fax to 781-893-7957. Questions can be directed to Inside Sales Manager at 781-906-7952.
- Quantity in commerce
- 2480 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Israel, Jordan, Russia, Saudi Arabia, South Africa, Turkey, Iraq, Lebanon, Morocco, Tunisia, Syria, India, Argentina, Australia, Austria, New Zealand, China, UK, Germany, Russia, Belgium, Italy, France, Sweden, and Ireland.
- Product codes and lots
- Lot #: 0006986, Exp Date: 2015-01; Lot #: 0048865, Exp. Date: 2015-02; Lot #: 0211585, Exp. Date: 2015-08; Lot #: 0244725, Exp. Date 2015-09