Z-0747-2011
HiSpeed DX/I (2201000-2)
GE Healthcare / initiated 2010-02-01 / Terminated / root cause: software
GE Healthcare, LLC began a recall of HiSpeed DX/I (2201000-2) on . The reason FDA records is "A defect with the hardware and software displaying the incorrect dose data". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0747-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- HiSpeed DX/I (2201000-2)
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- A defect with the hardware and software displaying the incorrect dose data.
- Root cause tag
- software
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Quantity in commerce
- 102
- Distribution
- Nationwide Distribution
- Product codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.