Gendex Dental Systems, Gendex eXpert DC Intraoral X-Ray System; a dental x-ray system; Model 110-0205G1...
Gendex Dental Systems began a recall of Gendex Dental Systems, Gendex eXpert DC Intraoral X-Ray System; a dental x-ray system; Model 110-0205G1, catalog number EXPERTDC75NA on . The reason FDA records is "The intraoral x-ray systems were installed improperly with incorrect screws which allow the unit to pull away from the wall when the arm is fully extended. This condition may harm the patient". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1248-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Gendex Corp.
- Product
- Gendex Dental Systems, Gendex eXpert DC Intraoral X-Ray System; a dental x-ray system; Model 110-0205G1, catalog number EXPERTDC75NA
- Product code
- EHD Unit, X-Ray, Extraoral With Timer
- Reason for recall
- The intraoral x-ray systems were installed improperly with incorrect screws which allow the unit to pull away from the wall when the arm is fully extended. This condition may harm the patient.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Gendex sent a Urgent Medical Device Field Correction letter dated December 17, 2010 to the consignee, informing them that the x-ray units were installed with improper sized lag bolts and requesting that they contact Gendex Customer Service at 1-800-323-8029 to complete the field correction.
- Quantity in commerce
- 5 units
- Distribution
- California
- Product codes and lots
- Model 110-0205G1, catalog number EXPERTDC75NA; serial numbers 1987591 - 09/2010, 1987597 - 09/2010, 1987598 - 09/2010, 1987599 - 09/2010 and 1987600 - 09/2010