Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0954-2011

Covidien, LigaSure, Blunt Tip Laparoscopic Sealer/Divider, 5 mm-37 cm, REF LF1537, Sterile EO, Covidien...

Medtronic, Inc. / initiated 2010-12-08 / Terminated / root cause: design

Covidien LP began a recall of Covidien, LigaSure, Blunt Tip Laparoscopic Sealer/Divider, 5 mm-37 cm, REF LF1537, Sterile EO, Covidien, Valleylab, a division of Tyco Healthcare Group LP... on . The reason FDA records is "During surgical procedures, the device may energize when clamping without pushing the activation button, which could result in unintended energy delivery to tissue". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0954-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Covidien, LigaSure, Blunt Tip Laparoscopic Sealer/Divider, 5 mm-37 cm, REF LF1537, Sterile EO, Covidien, Valleylab, a division of Tyco Healthcare Group LP, 5920 Longbow Drive, Boulder, CO 80301 Intended use: LigaSure 5 mm devices are electrosurgical instruments designed to permanently seal and divide vessels, tissue bundles, and lymphatics.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
During surgical procedures, the device may energize when clamping without pushing the activation button, which could result in unintended energy delivery to tissue.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Covidien, sent a "URGENT: PRODUCT RECALL" letter dated December 6, 2010, to the customers. The letter described the product, problem and the action to be taken by the customers. The customers were instructed to remove product from inventory, immediately quarantine the product and return any remaining stock of the lots listed to: Covidien Energy-based devices, Attention: R82, Building 4/Receiving, 5920 Longbow Drive, Boulder, CO 80301. The customers were also instructed to complete and return the Customer Response Form immediately via mail (in the envelope provided) or fax to (303) 516-8307. To return any identified product for credit or replacement, contact Customer Service at 1-800-722-8772 Option 1. Should you have any medical questions concerning this recall, please contact your local sales representative or the Covidien Clinical information Hotline at (800) 255-8522 (option 1).
Quantity in commerce
25,009 units
Distribution
Worldwide distribution: USA and countries including: Australia, Canada, France, Germany, Japan, Puerto Rico, Singapore, and Uruguay.
Product codes and lots
Lot No.: 188856, 190048, 190107, and 190307 through 193333 (88 lots).