ThermiSense model 5700-A5 The ThermiSense Model 5700-A5 is a nasal cannula that detects thermal variation...
Salter Laboratories, Division of Regulatory Affairs began a recall of ThermiSense model 5700-A5 The ThermiSense Model 5700-A5 is a nasal cannula that detects thermal variation during breathing cycles and transmits those... on . The reason FDA records is "The incorrect length and connector were utilized in lot 110810 of the ThermiSense Alice 5, making the product unusable as it will not interface with the sleep labs equipment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1028-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Salter Labs
- Product
- ThermiSense model 5700-A5 The ThermiSense Model 5700-A5 is a nasal cannula that detects thermal variation during breathing cycles and transmits those variations to a remote Alice-5 Polysomnograph via electrical connectors to monitors an adult patient's breathing rate in a sleep laboratory study
- Product code
- BZQ Monitor, Breathing Frequency
- Reason for recall
- The incorrect length and connector were utilized in lot 110810 of the ThermiSense Alice 5, making the product unusable as it will not interface with the sleep labs equipment.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Salter Laboratories, Inc. sent an URGENT PRODUCT RECALL LETTER dated December 20, 2010, to all affected customers. The letter identfied the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory for this product and lot number. Contact Salter to arrange for the return of the product. Complete the attached reply form, confirming receipt of the letter, and fax it back to Salter Labs at 1-800-628-4690. For questions regarding this recall call 661-854-3166.
- Quantity in commerce
- 41 units
- Distribution
- Nationwide Distribution including CA, CO, DE, ID, IN, KY, MD, MO, and OH,
- Product codes and lots
- Lot 110810
FIRST ARTICLE / WEEKLY