Z-0993-2011
Arstasls One Access System; Models: AAD100, AAD101, Part Numbers: FG-02279, FG-03010; The system consists of...
Arstasis, Inc. / initiated 2010-10-19 / Terminated / root cause: design
Arstasis, Inc began a recall of Arstasls One Access System; Models: AAD100, AAD101, Part Numbers: FG-02279, FG-03010; The system consists of the following components: (1) Device... on . The reason FDA records is "Device has the potential to fracture and sheath separation during use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0993-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arstasis, Inc.
- Product
- Arstasls One Access System; Models: AAD100, AAD101, Part Numbers: FG-02279, FG-03010; The system consists of the following components: (1) Device Sheath/Anchor, (2) Shaft and (3) Handle with control features. Product is manufactured and distributed by Arstasis Inc., San Carlos, CA The Arstasis One Access System is a sterile, single use system intended to provide access for the percutaneous introduction of devices into the peripheral vasculature and to promote hemostasis at the arteriotomy site as an adjunct to manual compression. The System is Indicated for use in patients undergoing diagnostic femoral artery catheterization procedures using SF or 6F Introducer sheaths.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Device has the potential to fracture and sheath separation during use.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, Arstasis, initiated a market withdrawal and sent notification via letters dated October 19, 2010, to customers informing the customers that Arstasis is implementing a product improvement and will replace all existing inventory that the customer has on their shelf with the improved product. The customers were instructed to work closely with their local Arstasis territory manager to ensure that their product replacement is handled satisfactorily, and complete and return the PRODUCT REPLACEMENT and PRODUCT RETURN forms to Arstasis Customer Service via fax at 1-877-688-8417 or email to [email protected]. If you have any questions or concerns, contact Customer Service toll free at 1-877-594-4545.
- Quantity in commerce
- 330 units
- Distribution
- Worldwide Distribution
- Product codes and lots
- Lot Numbers: 09I10268, 1OC26337, 09J06281, 10C12334