Z-0992-2011

Steerable Coronary Sinus Catheter Kit, REF SCS01, 9 Fr x 20.4", Contains Phthalates, Sterile EO, Edwards...

Edwards Lifesciences LLC / initiated 2010-12-03 / Terminated / root cause: design

Edwards Lifesciences Llc began a recall of Steerable Coronary Sinus Catheter Kit, REF SCS01, 9 Fr x 20.4", Contains Phthalates, Sterile EO, Edwards Lifesciences LLC, Irvine, CA 92614. The Steerable... on . The reason FDA records is "Defective product may result in the silicone valve becoming dislodged from the introducer which can result in injury". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0992-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Steerable Coronary Sinus Catheter Kit, REF SCS01, 9 Fr x 20.4", Contains Phthalates, Sterile EO, Edwards Lifesciences LLC, Irvine, CA 92614. The Steerable Coronary Sinus Catheter is a three-lumen, 9 Fr catheter with an elastomeric balloon near its tip for occluding the coronary sinus in order to retrograde perfuse the coronary circulation. An 11 Fr catheter introducer sheath, which is compatible with the Steerable Coronary Sinus Catheter contamination guard, is packaged with these catheters.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Defective product may result in the silicone valve becoming dislodged from the introducer which can result in injury.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Edwards Lifesciences, sent an "URGENT PRODUCT RECALL - ACTION REQUIRED" letter dated December 3, 2010 to customers. The letter describes the product, problem and actions to be taken by the customers. The customers were instructed to check their inventory and return the unused devices; complete and return the INTROCSC Introducer Recall Response form to the attention of Customer Service, Edwards LifeSciences, One Edwards Way, Irvine, CA at 949-250-3489. If you have any questions, please contact Edwards Customer Service at 800-424-3278 from 8:00am - 4:30pm Pacific Time.
Quantity in commerce
90 kits
Distribution
Nationwide Distribution
Product codes and lots
Lot No.: 666556, 669235, 678238, 751043, 755821, 763122