Z-1063-2011
Inrad sample master aspiration biopsy needle with hiliter sharp bevel needle, 22 ga x 2 3/8'' (6 cm), model...
RLS Interventional Inc. dba Inrad / initiated 2010-12-20 / Terminated / root cause: manufacturing process
Inrad, Inc began a recall of Inrad sample master aspiration biopsy needle with hiliter sharp bevel needle, 22 ga x 2 3/8'' (6 cm), model 41222, Inrad, Inc. Kentwood, MI. The device is... on . The reason FDA records is "This lot of product has the wrong expiration date on the carton label. The pouches contained the correct expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1063-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RLS Interventional Inc. dba Inrad
- Product
- Inrad sample master aspiration biopsy needle with hiliter sharp bevel needle, 22 ga x 2 3/8'' (6 cm), model 41222, Inrad, Inc. Kentwood, MI. The device is intended for use during percutaneous biopsy and drainage procedures to obtain tissues or fluid specimens.
- Product code
- GAA Needle, Aspiration And Injection, Disposable
- Reason for recall
- This lot of product has the wrong expiration date on the carton label. The pouches contained the correct expiration date.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The firm, Inrad Inc., sent an "URGENT FIELD CORRECTION" notice dated December 20, 2010 to its customers. The notice described the product, problem and actions to be taken by the customers. The customers were instructed to immediately examine their inventory for the affected product. If the recalled product was in stock, the customers were told to removed the affected product from the carton and dispose of carton; to disregard carton label expiration date and reference the expiration date appearing on the pouch label only, and to complete, sign and return the attached "Carton Expiration Date" INFORMATION FORM via fax to 616-301-779 or mail to Inrad Inc., ATTN: Director RA/QA, 4375 Donker Court SE, Kentwood, MI 49512. If you have any additional questions about this action, please call 800-558-4647 extension 102.
- Quantity in commerce
- 250
- Distribution
- Nationwide Distribution: USA including states of: FL, GA, KY, NH, NJ, NM, PA , SC, WA, and WI.
- Product codes and lots
- 6100-7001.