Z-0941-2011

Prelude Short Sheath Introducer, 7F-SMT, 4 cm, REF No: PSS-7F-4-038MT, Sterile EO, Merit Medical, South...

Merit Medical Systems, Inc. / initiated 2010-12-17 / Terminated

Merit Medical Systems, Inc. began a recall of Prelude Short Sheath Introducer, 7F-SMT, 4 cm, REF No: PSS-7F-4-038MT, Sterile EO, Merit Medical, South Jordan, Utah. Intended to provide access and facilitate... on . The reason FDA records is "Introducer tips may detach and embolize during use, or the tip material may elicit a thrombogenic response". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0941-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Prelude Short Sheath Introducer, 7F-SMT, 4 cm, REF No: PSS-7F-4-038MT, Sterile EO, Merit Medical, South Jordan, Utah. Intended to provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. Can also provide access to a native or synthetic graft used for hemodialysis.
Product code
DYB Introducer, Catheter
Reason for recall
Introducer tips may detach and embolize during use, or the tip material may elicit a thrombogenic response.
FDA root cause
Component change control
Action
Merit Medical Systems, Inc., issued an "Urgent Product Recall Notice" dated December 17, 2010, and contacted all customers (sales reps and authorized distributors) via email, phone and/or fax. The notice described the product, problem and action to be taken by the customers (Sales reps/Distributors). The customers were instructed to contact all their customers to identify, quarantine and return all unused inventory and extend the recall to any of their customers who may have received the affected products. The customers were instructed to complete and return the Manufacturer RECALL NOTIFICATION and Product Retrieval Form via fax to 1-801-208-3378 and send original to: the Manager, Customer Service, Merit Medical Systems, Inc., 1600 West Merit Parkway, South Jordan, UT 84095-2416. Questions should be directed to the Senior Product Manager at 801-208-4344 or the Manager, Customer Service at 801-208-4365.
Quantity in commerce
378 units
Distribution
Nationwide Distribution
Product codes and lots
Lot Number: H179575