Z-1518-2011

LICOX CMP Tissue Oxygen Monitor; Catalog numbers REF AC31. Manufacturer GMS Gesellschaft fur medizinische...

Integra LifeSciences Corporation / initiated 2010-12-17 / Terminated / root cause: manufacturing process

Integra LifeSciences Corp. began a recall of LICOX CMP Tissue Oxygen Monitor; Catalog numbers REF AC31. Manufacturer GMS Gesellschaft fur medizinische, Sondentechnik mbH, Dorfstrasse 2, 24247 Mielkendorf... on . The reason FDA records is "Certain LICOX CMP Tissue Oxygen Monitors could allow for underestimation of temperature values". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1518-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
LICOX CMP Tissue Oxygen Monitor; Catalog numbers REF AC31. Manufacturer GMS Gesellschaft fur medizinische, Sondentechnik mbH, Dorfstrasse 2, 24247 Mielkendorf, Germany; Distributed by Integra NeuroSciences, 311 Enterprise Drive, Plainsboro, New Jersey 08536 USA. Used for continuous determination of oxygen partial pressure in cerebrospinal fluid and brain tissue.
Product code
GWM Device, Monitoring, Intracranial Pressure
Reason for recall
Certain LICOX CMP Tissue Oxygen Monitors could allow for underestimation of temperature values.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Integra Life Sciences notified their Integra Neurospecialist via email and their customers via Fed Ex overnight mail on December 17, 2010 with an Urgent: Product Recall letter. The letter identified the affected product and the reason for the recall. The letter states that an Integra Neurospecialist will contact customers to arrange for the return of their Licox monitor to the Service Center for recalibration. A loaner will be provided until the recalibration is complete. The attached "Recall Acknowledgement and Return Form" must be completed and returned to Integra. If there are any additional questions, customers should contact Gary Rabinovich at 609-936-6864.
Quantity in commerce
26 units
Distribution
Worldwide Distribution
Product codes and lots
Lot/Serial numbers: 1982, 1988, 1990, 1991, 1992, 1993, 1994, 1995, 1996, 1997, 1998, 1999, 2000, 2001, 2003, 2005, 2007, 2008, 2009, 2010, 2011, 2012, 2014, 2046, 2047, 2051, and 2053.