Z-1042-2011

Vertex Max Drill Tube, Catalog Number 9733140, a component of the Navigated Vertex Max Drill Guide Kit...

Medtronic Navigation, Inc / initiated 2010-12-15 / Terminated / root cause: design

Medtronic Navigation, Inc began a recall of Vertex Max Drill Tube, Catalog Number 9733140, a component of the Navigated Vertex Max Drill Guide Kit, Catalog Number 9733370. Medtronic, Inc., Louisville... on . The reason FDA records is "Incorrectly manufactured drill tube could cause incorrect screw placement during spinal surgery". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1042-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Vertex Max Drill Tube, Catalog Number 9733140, a component of the Navigated Vertex Max Drill Guide Kit, Catalog Number 9733370. Medtronic, Inc., Louisville, CO. Used in conjunction with the StealthStation navigation systems to assist in screw placement during spine surgery.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Incorrectly manufactured drill tube could cause incorrect screw placement during spinal surgery.
Root cause tag
design
FDA root cause
Component design/selection
Action
Consignees were notified by a Product Correction Notification letter on 12/15/2010 and given the affected product and a description of the problem. The letter stated that a Medtronic representative would contact customers to visit their facility and replace the affected guide with one that has been correctly manufactured. Additional information is available at 800-595-9709.
Quantity in commerce
15 devices
Distribution
Worldwide Distribution -- USA, including states of GA, IL, IN, MN, MO, OH, OR, PA, and TX and countries of Denmark, Finland, Singapore, Spain, and UK.
Product codes and lots
Lot number: 100721