COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD
Roche Molecular Systems, Inc. began a recall of COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD on . The reason FDA records is "There is an issue regarding liquid level detection of the CS1 Magnetic Glass Particles (MGP) Reagent Cassette of the COBAS AmpliPrep /COBAS TaqMan HBV v2.0 Test on the COBAS AmpliPrep instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1459-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Molecular Systems Inc
- Product
- COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD
- Product code
- MKT Hepatitis Viral B Dna Detection
- Reason for recall
- There is an issue regarding liquid level detection of the CS1 Magnetic Glass Particles (MGP) Reagent Cassette of the COBAS AmpliPrep /COBAS TaqMan HBV v2.0 Test on the COBAS AmpliPrep instrument.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Roche sent Urgent Medical Device Correction Letters to their consignees on December 28, 2010 via Fed Ex. The letters advise the customers to verify the correct configuration of the CSI reagent cassette, remind customers to follow the package insert instructions for including all required controls for each batch and advise that review of previously generated results may be warranted. A UMDC Faxback form was attached for the consignee to complete. For questions regarding this recall call 908-253-7569.
- Quantity in commerce
- 93 kits to US; 22,119 kits to foreign
- Distribution
- Worldwide Distribution - USA including CA, CT, IN, MI, MN, NC, NE, NY, OR, TX, WA, and WI and the countries of Australia, Austria, Azerbaijan, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Egypt, Finland, France, Germany, Ghana, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mauritania, Mongolia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Singapore, Slovakia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, and the United Kingdom
- Product codes and lots
- Catalog number 05027012190, 05393850190, 05393850190, 05393868190 Lot #K11236, K15440, L00846, M05135, M11190, M12625, M14232, N00159, N02341, N05553, N05554, N07543, N08567