Z-1459-2011

COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD

Roche Molecular Systems Inc / initiated 2010-12-28 / Terminated / root cause: manufacturing process

Roche Molecular Systems, Inc. began a recall of COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD on . The reason FDA records is "There is an issue regarding liquid level detection of the CS1 Magnetic Glass Particles (MGP) Reagent Cassette of the COBAS AmpliPrep /COBAS TaqMan HBV v2.0 Test on the COBAS AmpliPrep instrument". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1459-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Molecular Systems Inc
Product
COBAS AmpliPrep/COBAS TaqMan HBV v2.0 US-IVD; COBAS AmpliPrep/COBAS TaqMan HBV v2.0 CE-IVD
Product code
MKT Hepatitis Viral B Dna Detection
Reason for recall
There is an issue regarding liquid level detection of the CS1 Magnetic Glass Particles (MGP) Reagent Cassette of the COBAS AmpliPrep /COBAS TaqMan HBV v2.0 Test on the COBAS AmpliPrep instrument.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Roche sent Urgent Medical Device Correction Letters to their consignees on December 28, 2010 via Fed Ex. The letters advise the customers to verify the correct configuration of the CSI reagent cassette, remind customers to follow the package insert instructions for including all required controls for each batch and advise that review of previously generated results may be warranted. A UMDC Faxback form was attached for the consignee to complete. For questions regarding this recall call 908-253-7569.
Quantity in commerce
93 kits to US; 22,119 kits to foreign
Distribution
Worldwide Distribution - USA including CA, CT, IN, MI, MN, NC, NE, NY, OR, TX, WA, and WI and the countries of Australia, Austria, Azerbaijan, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Egypt, Finland, France, Germany, Ghana, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mauritania, Mongolia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Singapore, Slovakia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, and the United Kingdom
Product codes and lots
Catalog number 05027012190, 05393850190, 05393850190, 05393868190 Lot #K11236, K15440, L00846, M05135, M11190, M12625, M14232, N00159, N02341, N05553, N05554, N07543, N08567