The Panorama HFO system has the Traceable Item Identification (12NC): 9896 030 17741-SW R 2.5.1 (shipped as R...
Philips Medical Systems North America Co. Phillips began a recall of The Panorama HFO system has the Traceable Item Identification (12NC): 9896 030 17741-SW R 2.5.1 (shipped as R 2.5.1, upgraded to R 2.5.3 via FCO). 9896 030... on . The reason FDA records is "In the ExamCard MobiFlex the Patient Position is "Prone" which must be "Supine". The result is mislabeling of left and right side of the patient. If this is not detected by the operator there is a risk of wrong...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1458-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- The Panorama HFO system has the Traceable Item Identification (12NC): 9896 030 17741-SW R 2.5.1 (shipped as R 2.5.1, upgraded to R 2.5.3 via FCO). 9896 030 19641-SW R 2.5.3 9896 030 19861-SW R 2.6.1 9896 030 19862-SW R 2.6.1 9896 030 19863-SW R 2.6.3
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- In the ExamCard MobiFlex the Patient Position is "Prone" which must be "Supine". The result is mislabeling of left and right side of the patient. If this is not detected by the operator there is a risk of wrong treatment or the need of additional X-RAY during treatment.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Philips began sendng out the URGENT-Device Correction Notice letter dated February 5, 2010, ro all affected consignees on February 10, 2010. The consignees were advised to change the patient position from "Prone" immediately to "Supine" and store the corrected ExamCard in the hospital database if they use the ExamCard, MobiFlex dual Injection. Consignees were instructed to call Philips Healthcare Call Center at 800-722-9377 #5 (then follow prompts) and also reference "FCO 78100318" with any communication.
- Quantity in commerce
- 82 units
- Distribution
- Nationwide Distribution including the following states: AL, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NE, NH, NJ, NY, OH, OR, SC, TN, TX, VA, VT, WI, and WV.AL, CA, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NE, NH, NJ, NY, OH, OR, SC, TN, TX, VA, VT, WI, and WV. There was no foreign consignees.
- Product codes and lots
- Site Numbers: 533860, 533900, 535230, 535664, 538574, 539374, 540482, 541065, 541851, 542236, 544039, 544148, 544696, 545034, 545732, 545981, 546206, 547843, 547897, 548040, 548064, 549389, 549909, 550324, 550439, 551025, 551583, 552453, 552526, 552609, 553116, 554715, 556146, 556161, 556895, 557599, 41443708, 41444456, 41444458, 41444745, 41445015, 41445018, 41445530, 41445594, 41445595, 41446119, 41455905, 41455908, 41455913, 41455915, 41455923, 41455938, 41455940, 41455954, 42047935, 42057444, 42293607, 42392891, 42608348, 42850814, 42892497, 42917294, 43708886, 44062048, 44247559, 44605983, 45130117, 45245963, 45246494, 45246504, 45445404, 47157157, 47230785, 47455618, 47480028, 47848863, 49060365, 49060404, 49416204, 49424702, 49776644, and 49861074.