Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device Product Code STRAP25. Intended for fixation of...
Ethicon, Inc. began a recall of Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device Product Code STRAP25. Intended for fixation of prosthetic material to soft tissues in various... on . The reason FDA records is "The Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device may have the potential for the sterile barrier to be compromised due to a defect in the packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1260-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon, Inc
- Product
- Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device Product Code STRAP25. Intended for fixation of prosthetic material to soft tissues in various minimally invasive and open surgical procedures, such as hernia repairs.
- Product code
- GDW Staple, Implantable
- Reason for recall
- The Ethicon SecureStrap 5 mm Absorbable Strap Fixation Device may have the potential for the sterile barrier to be compromised due to a defect in the packaging.
- Root cause tag
- packaging
- FDA root cause
- Pending
- Action
- Ethicon sent Urgent Voluntary Product Recall letters dated, December 28, 2010, to their customers on December 29, 2010 via UPS. The letter identified the affected product and described the reason for recall. Customers are to discontinue use and sampling immediately. They are to return the product using the enclosed prepaid shipping label along with the enclosed business reply card regardless of whether they have any affected product or not. Customers should contact 1-877-551-7147 or speak to their manager if they have questions regarding this recall.
- Quantity in commerce
- 4,110
- Distribution
- Worldwide Distribution -- USA and Canada.
- Product codes and lots
- All Lots. Batch Numbers: CJM838,CJZ014, CJZ065, CJZ228, CJZ456, CJZ542, CJZ625, CJ6116, CJ6293, CJ6402, CJ6572, CJ6852, CJK043, CJK216, CJK330, CJK536, CJK659, CKM791, CKZ068, CKZ437, CKZ652, and CK6002.