ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; ATG45...
Ethicon Endo-Surgery Inc began a recall of ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; ATG45; ATB45; ATW45; ATS45NK; and NAW45. Intended for... on . The reason FDA records is "The firm initiated this second subsequent recall because the affected devices can remain closed and clamped on tissue during surgical procedures due to the fact that the mechanism which connects the articulation joint...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1068-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon Endo-Surgery Inc
- Product
- ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Product Codes: ATS45; LONG45A; 6TB45; ATG45; ATB45; ATW45; ATS45NK; and NAW45. Intended for transection, resection, and/or creation of anastomoses.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- The firm initiated this second subsequent recall because the affected devices can remain closed and clamped on tissue during surgical procedures due to the fact that the mechanism which connects the articulation joint to the device shaft is insufficient.
- FDA root cause
- Other
- Action
- Ethicon Endo-Surgery issued "Urgent Expanded Device Recall Event 1981" notification letters dated April 7, 2009, to their customers. They provided a list of all Affected Product Codes of the ETS FLEX45 Articulating Endoscopic Linear Cutters (including the Product Code: ATS45 from the 1st recall) and Affected Flex Tray Codes subject to recall. The notification document reminds the Operating Room and Materials Management customers of the problem which necessitated the previous recall (Z-0361-2011). It goes on to instruct their customers to immediately discontinue use of the affected products. They are to complete and fax back the Business Reply Form, Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle, Replacement product should be received within 30-60 days after return, Customers can contact Stericycle concerning this recall at 1-888-365-4680, reference Event 1981.
- Quantity in commerce
- 29,403 units
- Distribution
- Nationwide Distribution -- USA including the following states, AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MO, MS, MT, NC, NE, NH, NJ, MN, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, and Puerto Rico.
- Product codes and lots
- The recalled Lot Numbers are: E4MJ47, E4MF6W, E4MK84, E4MM23, E4MM64, E4MP4A, E4MR15, E4MT07, E4MT4Y, F4MU2R, F4MY91, E4MG53, E4MG5K, E4MG5L, E4MN39, E4MG89, E4MG9M, E4MH2L, E4ML6J, E4MG77, E4MG8A, E4MH18, E4MJ3K, F4MX8C, and F4N799.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
