Philips FloTrak Elite modules: FloTrak Elite P/N 1044971 FloTrak Elite P/N 1051005 NM3 is a multi-parameter...
Respironics Novametrix, LLC. began a recall of Philips FloTrak Elite modules: FloTrak Elite P/N 1044971 FloTrak Elite P/N 1051005 NM3 is a multi-parameter patient monitor that utilizes the FloTrak Elite... on . The reason FDA records is "Software: two displayed respiratory parameters are incorrect due to errors in the system software: MValv (alveolar minute ventilation) and Vt/kg (tidal volume/patient weight). In both instances, the displayed parameters...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2168-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Respironics Novametrix, LLC.
- Product
- Philips FloTrak Elite modules: FloTrak Elite P/N 1044971 FloTrak Elite P/N 1051005 NM3 is a multi-parameter patient monitor that utilizes the FloTrak Elite module to perform some of the calculations.
- Reason for recall
- Software: two displayed respiratory parameters are incorrect due to errors in the system software: MValv (alveolar minute ventilation) and Vt/kg (tidal volume/patient weight). In both instances, the displayed parameters are higher than actual.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Respironics issued Field Action Notification Letter via Fed'x on 1/4/11, notifying customers of the issue and to instruct them on conditions for continued use and return of the device. The device will be updated with revised software to correct the calculation errors. Any questions or need additional information about this recall action, Contact Respironics US Customer Service at 1-800-345-6443 for customers residing in the USA or +011 (203)697-6488 for International customers.
- Quantity in commerce
- 38 units
- Distribution
- Nationwide Foreign: Belgium, China, Columbia, Egypt, England, India, Italy, Saudi Arabia, and Spain
- Product codes and lots
- Serial Numbers: 167-000475 167-000496 175 235 236 237 238 239 240 356 376 389 414 483 502 510 511 686 693 700 723 724 725 612 146 155 171 173 and 178