Z-1010-2011

Fraxel re:store Dual Laser System----a laser surgical instrument for use in general and plastic surgery and...

Solta Medical, Inc. / initiated 2010-12-15 / Terminated / root cause: software

Solta Medical, Inc. began a recall of Fraxel re:store Dual Laser System----a laser surgical instrument for use in general and plastic surgery and in dermatology. Model Numbers: MC-SYS-SR1500-D-US... on . The reason FDA records is "The product has the potential to cause inadvertent laser firing which may lead to patient/operator injury if the problem recurs". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1010-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Solta Medical, Inc.
Product
Fraxel re:store Dual Laser System----a laser surgical instrument for use in general and plastic surgery and in dermatology. Model Numbers: MC-SYS-SR1500-D-US; MC-SYS-SR1500-D-US-LOANER; MC-SYS-SR1500-D-UPG-US; MC-SYS-SR 1500-0-1 NTL; MC-SYS-SR1500-D-I-LOANER; MC-SYS-SR1500-D-UPG-INTL; Product is manufactured and distributed by Solta Medical, Inc., Hayward, CA The Fraxel re:store Dual Laser System is indicated for use in dermatological procedures requiring the coagulation of soft tissue.
Product code
ONG Powered Laser Surgical Instrument With MicrobeamFractional Output
Reason for recall
The product has the potential to cause inadvertent laser firing which may lead to patient/operator injury if the problem recurs.
Root cause tag
software
FDA root cause
Software design
Action
The firm, Solta Medical, issued three "DEVICE CORRECTION" letters, one dated December 15, 2010, the second one date December 17, 2010 and the third one dated December 20, 2010, to all affected customers. The letters described the product, problem and action to be taken by the customers. The customers were instructed to follow the instructions carefully, sign and date the bottom of the instruction form upon completion of the update and promptly return the signed form and USB stick in the enclosed, return-addressed envelope (Product Support, Solta Medical, 25881 Industrial Blvd., Hayward, CA 94545) Note: Solta Medical sales personnel will assist with the software update process so, if your local representative has already updated your system, you do not need to do anything further. If you have any questions, please call Solta Product Support at (510) 259-7291.
Quantity in commerce
352 units
Distribution
Worldwide Distribution: USA and countries including: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Egypt, Germany, Greece, Hong Kong, India, Italy, Japan, Korea, Kuwait, Mexico, Netherlands, New Zealand, Pakistan, Philippines, Poland, Russia, Singapore, South Africa, Spain, Thailand, Turkey, Ukraine, United Kingdom, and Vietnam
Product codes and lots
Serial Numbers: F1196J, J0007, J0010-J0058, J0060-J0203, J0205, J0207-J0209, J0211-J0337, J0339-J0359, J0361-J0414, J0417-J0419, J0426