Z-1848-2011

HEMOCUE AB GLUCOSE 201 DM ANALYZER, in vitro diagnostic device for quantitative determination of blood...

Hemo Cue, Inc. / initiated 2010-10-07 / Terminated / root cause: manufacturing process

Hemo Cue, Inc. began a recall of HEMOCUE AB GLUCOSE 201 DM ANALYZER, in vitro diagnostic device for quantitative determination of blood glucose level on . The reason FDA records is "Incorrect configurations settings in software. As a result, certain error codes do not trigger appropriately. These error codes are: E01, E02 and E05 which are all related to the optronic performance of the analyzer...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1848-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hemo Cue, Inc.
Product
HEMOCUE AB GLUCOSE 201 DM ANALYZER, in vitro diagnostic device for quantitative determination of blood glucose level.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
Incorrect configurations settings in software. As a result, certain error codes do not trigger appropriately. These error codes are: E01, E02 and E05 which are all related to the optronic performance of the analyzer. Possible incorrect results.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Hemocue AB sent Customer Notification "Notice of Medical Device Correction on HemoCue Glucose 201 DM and HemoCue Hb 201 DM Analyzers" dated October 7, 2010, to customers notifying them of incorrect settings on error codes. The corrective action consists of an update of the configuration settings on the the affected HemoCue Glucose 201 DM and Hb 201 DM analyzers that are in the field. The affected units in the field will be corrected in accordance with a rework procedure that has been established by HemoCue AB.
Quantity in commerce
23
Distribution
Worldwide distribution, including USA, and Glucose 201 DM analyzers to Norway and the Netherlands; Hb 201 DM analzyers to the Netherlands, France, Ireland, Great Britain, Poland and Germany.
Product codes and lots
Hb Glucose 201 DM System: (USA): Serial Numbers:  0632720008; 0651720088; 0641720017; 0651720037, 0651720087 & 0651720092; 0707720047; 0639720004; 0641720004, 0641720008,& 0641710015.  Hb Glucose 201 DM System: (Euroope): Serial Numbers:   064172009; 0641720032; 0641720042, 0641720044 & 0644720026 & 0644720050; 0644720006 & 0651720002; 0644720022; 0644720039, 0644720044 & 0651720006.