HEMOCUE AB Hb 201 DM ANALYZER, in vitro diagnostic device for quantitative hemoglobin determination
Hemo Cue, Inc. began a recall of HEMOCUE AB Hb 201 DM ANALYZER, in vitro diagnostic device for quantitative hemoglobin determination on . The reason FDA records is "Incorrect configurations settings in software. As a result, certain error codes do not trigger appropriately. These error codes are: E01, E02 and E05 which are all related to the optronic performance of the analyzer...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1849-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hemo Cue, Inc.
- Product
- HEMOCUE AB Hb 201 DM ANALYZER, in vitro diagnostic device for quantitative hemoglobin determination.
- Product code
- GKR System, Hemoglobin, Automated
- Reason for recall
- Incorrect configurations settings in software. As a result, certain error codes do not trigger appropriately. These error codes are: E01, E02 and E05 which are all related to the optronic performance of the analyzer. Possible incorrect results.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Hemocue AB sent Customer Notification "Notice of Medical Device Correction on HemoCue Glucose 201 DM and HemoCue Hb 201 DM Analyzers" dated October 7, 2010, to customers notifying them of incorrect settings on error codes. The corrective action consists of an update of the configuration settings on the the affected HemoCue Glucose 201 DM and Hb 201 DM analyzers that are in the field. The affected units in the field will be corrected in accordance with a rework procedure that has been established by HemoCue AB.
- Quantity in commerce
- 79
- Distribution
- Worldwide distribution, including USA, and Glucose 201 DM analyzers to Norway and the Netherlands; Hb 201 DM analzyers to the Netherlands, France, Ireland, Great Britain, Poland and Germany.
- Product codes and lots
- HemoCue Hb 201 DM Analyzers, (USA): Serial Numbers: 0632620050; 0647620080, 0717620162 & 0717620185; 0650620105; 0717620156; 0717620169; 0717620186; 0647620088; 0717620170; 0717620157 & 0717620187; 0647620104, 0650620077, 0650620081, 0650620104, 0706620010, 0706620032, 0706620043, 0706620057, 0706620094, 0709620025 & 0709620031; 0650620066; 0650620112; 0717620152, 0717620153 & 0717620172; 0706620041; 0717620189; 0717620190; 0717620158; 0702620049; 0717620155; 0717620115; 0717620178 & 0717620179; 0717620159; 0717620065 & 0717620174; 0717620165; 0717620181; 0717620167; 0706620019 & 0706620020; 0717620171, 0717620173 & 0717620184; 0717620131; 0717620166, 0717620160, 0717620168 & 0717620180. HemoCue Hb 201 DM Analyzers, (Europe): Serial Numbers: 0630630060; 0639620042; 0640620003 & 0640620015; 0717620197; 0717620199; 0717620183; 0717620191; 0717620200; 0709620026 & 0709620121; 0709620050, 0709620097, 0717620193, 0717620194, 0717620195 & 0717620196; 0717620163 & 0717620164; 0709620034; 0709620023; 0717620198; 0647620069.