Z-1536-2011

Cylindrical Bullet Nose Reamer 12.5 mm x 250 mm, REF 475813, Biomet Orthopedics 66 East Bell Drive Warsaw IN...

Zimmer GmbH / initiated 2010-12-23 / Terminated / root cause: supplier

Biomet, Inc. began a recall of Cylindrical Bullet Nose Reamer 12.5 mm x 250 mm, REF 475813, Biomet Orthopedics 66 East Bell Drive Warsaw IN. The product is used to prepare the femoral canal... on . The reason FDA records is "The firm was notified by their supplier that during manufacture of this instrument, the 12 degree cutting edge was not created. Testing found that if the cutting edge is missing, the reamer will not perform as intended". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1536-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Cylindrical Bullet Nose Reamer 12.5 mm x 250 mm, REF 475813, Biomet Orthopedics 66 East Bell Drive Warsaw IN. The product is used to prepare the femoral canal for stem implantation.
Product code
HTO Reamer
Reason for recall
The firm was notified by their supplier that during manufacture of this instrument, the 12 degree cutting edge was not created. Testing found that if the cutting edge is missing, the reamer will not perform as intended.
Root cause tag
supplier
FDA root cause
Other
Action
The firm, BIOMET, sent an" URGENT MEDICAL DEVICE RECALL NOTICE" dated December 23, 2010 to its customers. The notice described the product, problem and actions to be taken by the customer. The customers were instructed to immediately locate, discontinue use of product and return the product to Biomet. Biomet distributors were told to notify hospital personnel of the recall if they had distributed the product to hospital customers. The Customers/Distributors were instructed to confirm receipt of the notice by calling 800-348-9500 extension 3755 and 3983 and to complete and return the Response Form included with the recall notice via fax to 574-372-1683. Questions related to this notice should be directed to 574-371-3755 or 574-372-3983, Monday through Friday, 8am to 5pm.
Quantity in commerce
51 all products
Distribution
Worldwide distribution: USA including states of: FL, OH, PA, IN, CA, and NJ; and countries of Argentina, Canada, Mexico, Netherlands and Spain.
Product codes and lots
060890, 077020 and 605240.