Z-0961-2011
SRD S5 Mast Roller Pump System 150, S5 Component, Part No.: 50-80-00, Made in Germany, Distributed by: Sorin...
LivaNova USA, Inc. / initiated 2010-11-12 / Terminated / root cause: design
Sorin Group USA, Inc. began a recall of SRD S5 Mast Roller Pump System 150, S5 Component, Part No.: 50-80-00, Made in Germany, Distributed by: Sorin Group USA, Inc., Arvada, CO 80004. Used to... on . The reason FDA records is "Touch screen may become unresponsive, inhibiting user input". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0961-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- SRD S5 Mast Roller Pump System 150, S5 Component, Part No.: 50-80-00, Made in Germany, Distributed by: Sorin Group USA, Inc., Arvada, CO 80004. Used to perform, control and monitor extracorporeal blood circulation during an operation for up to six hours.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- Touch screen may become unresponsive, inhibiting user input.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- All affected US customers and distributors were notified by certified mail on Nov 12, 2010 via a Field Correction letter. They were told that they would be contacted by Sorin to arrange an appointment to inspect and replace affected touch screens on site. They were told that they could safely continue using their S5 systems in accordance with the Operator's Manual and previously supplied instructions until the inspection and necessary replacement had been completed. A customer response form was also included for completion and return to the firm. Customers can contact the firm if they have any questions.
- Quantity in commerce
- 176 units
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Serial Numbers: 50K00313 to 50K00693
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
