Philips BV Endura, Rel. 2, 718074. Mobile C-Arm X-Ray systems offering Radiographic and Fluoroscopic...
Philips Medical Systems North America Co. Phillips began a recall of Philips BV Endura, Rel. 2, 718074. Mobile C-Arm X-Ray systems offering Radiographic and Fluoroscopic techniques in a wide variety of applications on . The reason FDA records is "The fixing strap of the belt holding the CBX Laser to the Image Intensifier may be detached during a procedure when used with the BV Libra and the BV Pulsera/Endura Systems". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1604-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips BV Endura, Rel. 2, 718074. Mobile C-Arm X-Ray systems offering Radiographic and Fluoroscopic techniques in a wide variety of applications.
- Product code
- IXL Device, Spot-Film
- Reason for recall
- The fixing strap of the belt holding the CBX Laser to the Image Intensifier may be detached during a procedure when used with the BV Libra and the BV Pulsera/Endura Systems.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 01/06/11, Philips sent an URGENT Device Correction Notice, dated 01/05/2011, to their consignees. The consignees were informed that the fixing strap of the Laser Aiming Device Image Intensifier may inadvertently detach when used with BV Libra, BV Endura Rel. 2, and BV Pulsera Rel 2.3 Mobile X-Ray systems. Specifically the draw latch of the fixing strap holding the CBX Laser to the Image Intensified can be accidentally released during a procedure when used with the above Mobile X-Ray systems. While waiting for Philips to implement their corrective action, the consignees are advised to fix the closed latch body of the draw latch to the metal band using a Tie Wrap. Without securing the latch with a Tie Wrap, the device should not be used. The consignees can contact Philips Call Center at 1-800-722-9377, select 5, and reference FCO 71800035 with any questions about this recall.
- Quantity in commerce
- 10 units
- Distribution
- Nationwide Distribution -- AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Washington, DC.
- Product codes and lots
- Site Numbers: 548686, 556612, 41416327, 41446303, 41566246, 41585083, 43896129, 45592395, 47198056, and 49736876.