Z-0939-2011

Linemaster Wireless Footswitch, VASER System, REF Vaser 2.0, Sound Surgical Technologies LLC, 357 South...

Solta Medical, Inc. / initiated 2010-12-21 / Terminated / root cause: manufacturing process

Sound Surgical Technologies, LLC began a recall of Linemaster Wireless Footswitch, VASER System, REF Vaser 2.0, Sound Surgical Technologies LLC, 357 South McCaslin Blvd., Suite 100, Louisville, CO 80027 on . The reason FDA records is "Surgical equipment, activated by a wireless foot switch, may remain powered on when switch is no longer depressed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0939-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Solta Medical, Inc.
Product
Linemaster Wireless Footswitch, VASER System, REF Vaser 2.0, Sound Surgical Technologies LLC, 357 South McCaslin Blvd., Suite 100, Louisville, CO 80027
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Surgical equipment, activated by a wireless foot switch, may remain powered on when switch is no longer depressed.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Customers were notified by e-mail with an Advisory Notice on 12/21/2010. The Notice identified the affected product and described the issue. The Notice also explained what the potential hazard involved along with a mechanism to eliminate the potential hazard. Sound Surgical would arrange a service call to implement the correction at the firm's expense. If customers experience this potential failure, they are to contact Sound Surgical Technologies at 1-888-677-8765 or 1-888-471-4777.
Quantity in commerce
27 units
Distribution
Nationwide Distribution -- AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, MA, MI, MO, NJ, NY, PA, and TX .
Product codes and lots
Serial numbers VNT10-0102 thru VNT10-0128.