Linemaster Wireless Footswitch, VASER System, REF Vaser 2.0, Sound Surgical Technologies LLC, 357 South...
Sound Surgical Technologies, LLC began a recall of Linemaster Wireless Footswitch, VASER System, REF Vaser 2.0, Sound Surgical Technologies LLC, 357 South McCaslin Blvd., Suite 100, Louisville, CO 80027 on . The reason FDA records is "Surgical equipment, activated by a wireless foot switch, may remain powered on when switch is no longer depressed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0939-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Solta Medical, Inc.
- Product
- Linemaster Wireless Footswitch, VASER System, REF Vaser 2.0, Sound Surgical Technologies LLC, 357 South McCaslin Blvd., Suite 100, Louisville, CO 80027
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Surgical equipment, activated by a wireless foot switch, may remain powered on when switch is no longer depressed.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Customers were notified by e-mail with an Advisory Notice on 12/21/2010. The Notice identified the affected product and described the issue. The Notice also explained what the potential hazard involved along with a mechanism to eliminate the potential hazard. Sound Surgical would arrange a service call to implement the correction at the firm's expense. If customers experience this potential failure, they are to contact Sound Surgical Technologies at 1-888-677-8765 or 1-888-471-4777.
- Quantity in commerce
- 27 units
- Distribution
- Nationwide Distribution -- AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, MA, MI, MO, NJ, NY, PA, and TX .
- Product codes and lots
- Serial numbers VNT10-0102 thru VNT10-0128.