Z-1226-2011
Thomas Medical Products, SafeSheath Long, (Tear-Away Sheath Introducer Set with Integral Hemostasis Valve)...
Merit Medical Systems, Inc. / initiated 2010-12-16 / Terminated / root cause: packaging
Thomas Medical Products Inc began a recall of Thomas Medical Products, SafeSheath Long, (Tear-Away Sheath Introducer Set with Integral Hemostasis Valve) for use during vascular interventional procedures... on . The reason FDA records is "Compromised sterility/weak or open seals". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1226-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Thomas Medical Products, SafeSheath Long, (Tear-Away Sheath Introducer Set with Integral Hemostasis Valve) for use during vascular interventional procedures. Part No. FCL-051-00, Catalog No. HLS2507MCN; and Part No. FCL-051-02, Catalog No. HLS2507MCN.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Compromised sterility/weak or open seals.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- GE Healthcare Thomas medical products Recall notification dated 16 December 2010 (GEHC Ref # 35003) was sent by letter via UPS Next Day Air and/or USPS certified mail.
- Quantity in commerce
- 28,679 for all products
- Distribution
- Distribution Nationwide USA and Germany.
- Product codes and lots
- Part No. FCL-051-00, Catalog No. HLS2507MCN, Lot #S33309; and Part No. FCL-051-02, Catalog No. HLS2507MCN, Lot #S33308.
FIRST ARTICLE / WEEKLY