Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used...

Elekta Limited / initiated 2007-08-23 / Terminated

Elekta, Inc. began a recall of Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used for radiation therapy treatments of malignant... on . The reason FDA records is "Under certain conditions it has been found that movement of the collimators can be induced during multi segment beams such as IMRT and Omniwedge which will not be detected by the Desktop Pro if high values are sent...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1555-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Desktop Pro 4.2, 5.0, 5.1, 6.0, 6.2 A user interface of rthe SL/SLi Series Linear Accelerators that are used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Under certain conditions it has been found that movement of the collimators can be induced during multi segment beams such as IMRT and Omniwedge which will not be detected by the Desktop Pro if high values are sent within the iCom tolerance table.
FDA root cause
Other
Action
The firm, Elekta, sent an "Important Notice A302" dated August 23, 2007 titled "Update to iCom Tables for Desktop Pro when used with Third Party R&V Systems",to all affected customers. On September 28, 2007, Elekta issued a second Important Notice A304 which reiterated the same information as was in Important Notice A302. The notices described the product, problem and actions to be taken. This notice provides a workaround on how to avoid the issue. In order to avoid this issues documented, Elekta recommended that the customers set the tolerance for collimators in the iCom tolerance table to be 3.0 degrees-no less than 3.0 and to use the Linac Record, if configured, as a means of checking the final collimator position. Elekta noted that the issue will be resolved in a future release of Desktop Pro. If you any questions related to this issue, please contact your local Elekta, third party R&V system representative or Elekta Limited at +44 (0)1293 654200.
Quantity in commerce
229 units
Distribution
Worldwide distribution: USA including Puerto Rico and countries including: Canada and Mexico.
Product codes and lots
134615, 134621, 134091, 136364, 134172, 136285, 134605, 134183, 136167, 136234, 134259, 134115, 134081, 136047, 136063, 134701, 134328, 134391, 134432, 134436, 134438, 136119, 134535, 134561, 134475, 134564, 134154, 134269, 134348, 134399, 136084, 136087, 136100, 134200, 134118, 134822, 134018, 136368, 134241, 134606, 134711, 134473, 134787, 134768, 134571, 134237, 134104, 136009, 134331, 134272, 136322, 136175, 136200, 134503, 136257, 134322, 134781, 134380, 134709, 136108, 134250, 134254, 134258, 134206, 134195, 134094, 134652, 134704, 136107, 134469, 134472, 134483, 134519, 134729, 136109, 134236, 134239, 134197, 134456, 134047, 134092, 134211, 136373, 134275, 134679, 136106, 136163, 134667, 136181, 134169, 134329, 134126, 134090, 134358, 134427, 134626, 134819, 134027, 134265, 134597, 136319, 134220, 134256, 134396, 134804, 134235, 136338, 134137, 134234, 134249, 136034, 136065, 136110, 136267, 136291, 134213, 134219, 136330, 134750, 136097, 136270, 134457, 134078, 134129, 136195, 134375, 134603, 136074, 134075, 134565, 134196, 134428, 134430, 134008, 134382, 134070, 134157, 134772, 136080, 134860, 134401, 134368, 134388, 134631, 134318, 136015, 136096, 134563, 136165, 136240, 134441, 134017, 134048, 134189, 134395, 134324, 134595, 134140, 134212, 136086, 134354, 136178, 136317, 136318, 134445, 136012, 136332, 134321, 136071, 136089, 136292, 136293, 134673, 134685, 136059, 136169, 134073, 134748, 134754, 134378, 134383, 134458, 134192, 134499, 136069, 136092, 134642, 134295, 134671, 136046, 136147, 134062, 136049, 136036, 136152, 136157, 136083, 136081, 134770, 136118, 136156, 136172, 134352, 134548, 134533, 136153, 136269, 136197, 136252, 136286, 136315, 136241, 134811, 134812, 136258, 136312, 134069, 136310, 134864, 134040, 134046, 134307