Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0966-2011

Cook Urological, Inc. Koyle Diaper Stent 6.0 Fr, Sterile, GPN G17667, REF KDSCF-060050

Cook Incorporated / initiated 2010-12-13 / Terminated / root cause: packaging

Vance Products Inc. dba Cook Urological Inc. & Cook OB/GYN began a recall of Cook Urological, Inc. Koyle Diaper Stent 6.0 Fr, Sterile, GPN G17667, REF KDSCF-060050 on . The reason FDA records is "The firm received a complaint that 15 units of the Koyle Diaper Stent were received in packages that had not been sealed. Each stent should be sealed in a poly-Tyvek pouch". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0966-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook Urological, Inc. Koyle Diaper Stent 6.0 Fr, Sterile, GPN G17667, REF KDSCF-060050.
Product code
KOD Catheter, Urological
Reason for recall
The firm received a complaint that 15 units of the Koyle Diaper Stent were received in packages that had not been sealed. Each stent should be sealed in a poly-Tyvek pouch.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
The firm sent an URGENT RECALL NOTICE to one consignee on 12/13/2010 (the other consignee returned all their product as a complaint). The letter informed the consignee of the recall and asked the customer to return the recalled product.
Quantity in commerce
25 stents
Distribution
CA, CT
Product codes and lots
U1973244