Z-1395-2011

SP Brand Strep A Cassette; Part Number: B1077-26; Lot #: STA0010037 The SP Brand Strep A Rapid Test Device is...

Alere San Diego, Inc. / initiated 2011-01-12 / Terminated

Biosite Inc Dba Innovacon Inc. began a recall of SP Brand Strep A Cassette; Part Number: B1077-26; Lot #: STA0010037 The SP Brand Strep A Rapid Test Device is a rapid chromatographic immunoassay for the... on . The reason FDA records is "The recall was initiated after the Investigation of customer-reported complaints, testing of market retention and returned product, It was determined that this lot of product has demonstrated an elevated Incidence of...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1395-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alere San Diego, Inc.
Product
SP Brand Strep A Cassette; Part Number: B1077-26; Lot #: STA0010037 The SP Brand Strep A Rapid Test Device is a rapid chromatographic immunoassay for the qualitative detection of Strep A antigen from throat swab specimens to aid in the diagnosis of Group A Streptococcal infection. This test is indicated for professional and point of care use only.
Product code
GTY Antigens, All Groups, Streptococcus Spp.
Reason for recall
The recall was initiated after the Investigation of customer-reported complaints, testing of market retention and returned product, It was determined that this lot of product has demonstrated an elevated Incidence of non-specific binding, potentially causing false positive results. A false positive test result could lead to the administration of antibiotics that are unnecessary. This exposes the
FDA root cause
Pending
Action
The firm, Alere San Diego, sent a "URGENT MEDICAL DEVICE RECALL" letter dated January 17, 2011. The letter described the product, problem and action to be taken by customers. The customers were instructed to: (1) Immediately discontinue use of the affected lot (if applicable) and inventory remaining full or partial kits at their site (if applicable); (2) Destroy any remaining inventory of this lot and document destruction per their site requirements (if applicable); (3) Complete and FAX the enclosed Verification Form within 10 days to confirm their receipt of this notice and to receive replacement of their remaining inventory of this lot (if applicable). The customers were ask to please return the form even if they have no remaining inventory or do not have any involved product in order to confirm receipt of this notification; and (4) to please ensure that all users of the device have received a copy of this Urgent Medical Device Recall. Customers with questions about the information contained in the notice were instructed to contact Technical Services at 888-246.7483, option 2 or by e-mail [email protected].
Quantity in commerce
814 kits/24,420 devices
Distribution
Nationwide distribution
Product codes and lots
Lot #: STA0010037