Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1128-2011

Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, Double Lumen...

Cook Incorporated / initiated 2010-10-04 / Terminated / root cause: manufacturing process

Cook, Inc. began a recall of Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, Double Lumen Polyurethane, 5.0 Fr12cm, C-UDLMY-501J- RSC-ABRM-HC-FST... on . The reason FDA records is "Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1128-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, Double Lumen Polyurethane, 5.0 Fr12cm, C-UDLMY-501J- RSC-ABRM-HC-FST, G44003, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
Quantity in commerce
51,473 total kits/trays
Distribution
Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
Product codes and lots
Part Number: C-UDLMY-501J-RSC- ABRM-HC-FST; Global Part Number G44003 lots 2238143, 2239207, 2240537, 2241784, 2252487, 2257544, 2267174, 2267995, 2273799, 2275424, 2276829, 2282636, 2286935, 2287725, 2290145, 2302073, 2311589, 2321689 and 2282636X.