Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1133-2011

Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, Triple Lumen...

Cook Incorporated / initiated 2010-10-04 / Terminated / root cause: manufacturing process

Cook, Inc. began a recall of Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, Triple Lumen Polyurethane, 5.0 Fr /15 cm, C-UTLMY-501J- LSC-ABRM-HC-FST... on . The reason FDA records is "Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1133-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Spectrum Central Venous Catheter Tray, Minocycline/Rifampin Antibiotic Impregnated, Triple Lumen Polyurethane, 5.0 Fr /15 cm, C-UTLMY-501J- LSC-ABRM-HC-FST, G44016, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
Product code
FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Reason for recall
Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
Quantity in commerce
51,473 total kits/trays
Distribution
Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
Product codes and lots
Part Number: C-UTLMY-501J-LLSC-ABRM-HC-FST; Global Part Number G44016, lots 2239666, 2242701, 2248940, 2253262, 2254312, 2255610, 2262114, 2263390, 2275431, 2280744, 2285512, 2309427, 2321454 and 2332349.