Z-1252-2011

Elekta Synergy XVI

Elekta Limited / initiated 2011-01-05 / Terminated / root cause: manufacturing process

Elekta, Inc. began a recall of Elekta Synergy XVI Intended use: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner on . The reason FDA records is "Flexmap calibration did not cover the full 360 degrees of gantry rotation (0 to 355 degrees) and that this caused an isocenter displacement with 2D PlanarView images required at 180 to -180". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1252-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Elekta Synergy XVI Intended use: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Flexmap calibration did not cover the full 360 degrees of gantry rotation (0 to 355 degrees) and that this caused an isocenter displacement with 2D PlanarView images required at 180 to -180.
Root cause tag
manufacturing process
FDA root cause
Other
Action
The firm, Elekta Limited, sent an "Important Notice A344" dated January 5, 2011 to all consignees/customers. The notice described the product, problem and action to be taken by the customers. The customers were instructed to follow any instructions or recommendations covered in the notice. The notice provides suggestions on how to check and prevent this situation from occurring; it instructs the User to file the document in the Important Notice section of the appropriate User Manual; to complete a FlexMap Check each time they acquire a new FlexMap and amend their local procedures to include the FlexMap Check and complete and return the Important Notice Confirmation of Receipt via Fax to: Field Support Administration, Elekta Limited, Linac House, Fleming Way Crawley, West Sussex, RH10 9RR UK +44 (0)1293 654401. If you have any questions, contact Elekta Limited at +44 (0)1293 654200 or www.elekta.com.
Quantity in commerce
223 units
Distribution
Nationwide distribution.
Product codes and lots
105257, 105803, 105941, 105982, 105984, 151024, 151038, 151051, 151053, 151055, 151073, 151074, 151084, 151110, 151130, 151141, 151151, 151156, 151157, 151160, 151168, 151173, 151178, 151216, 151245, 151250, 151256, 151258, 151259, 151260, 151298, 151301, 151309, 151317, 151319, 151326, 151327, 151328, 151340, 151352, 151359, 151360, 151362, 151363, 151371, 151387, 151391, 151394, 151396, 151398, 151410, 151411, 151412, 151416, 151418, 151423, 151432, 151433, 151435, 151441, 151443, 151446, 151467, 151479, 151504, 151507, 151514, 151515, 151517, 151519, 151521, 151522, 151523, 151530, 151538, 151540, 151545, 151557, 151558, 151566, 151567, 151568, 151574, 151584, 151585, 151587, 151595, 151598, 151601, 151611, 151616, 151623, 151626, 151627, 151628, 151632, 151634, 151636, 151640, 151653, 151664, 151672, 151674, 151675, 151678, 151683, 151684, 151689, 151694, 151695, 151696, 151698, 151699, 151700, 151704, 151705, 151708, 151710, 151714, 151722, 151724, 151728, 151729, 151734, 151736, 151751, 151759, 151760, 151763, 151764, 151765, 151770, 151774, 151779, 151782, 151783, 151785, 151789, 151791, 151793, 151801, 151802, 151803, 151805, 151808, 151809, 151810, 151811, 151812, 151820, 151828, 151839, 151842, 151847, 151848, 151855, 151860, 151864, 151866, 151867, 151875, 151881, 151882, 151885, 151886, 151892, 151931, 151933, 151942, 151952, 151953, 151955, 151956, 151958, 151959, 151960, 151961, 151965, 151969, 151977, 151978, 151981, 151989, 152008, 152019, 152023, 152031, 152043, 152051, 152064, 152079, 152080, 152091, 152098, 152103, 152115, 152116, 152117, 152123, 152126, 152154, 152158, 152163, 152174, 152175, 152176, 152177, 152192, 152200, 152210, 152211, 152214, 152217, 152220, 152222, 152223, 152232, 152246, 152257, 152271, 152301, 152307, 152308