MOSAIQ Oncology Information System
Siemens Medical Solutions USA, Inc. / initiated 2011-01-11 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of MOSAIQ Oncology Information System Intended Use : Electronic Patient Records Management on . The reason FDA records is "The potential for injury to a patient should the treatment data be incorrectly uploaded to patient records". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1572-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- MOSAIQ Oncology Information System Intended Use : Electronic Patient Records Management
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- The potential for injury to a patient should the treatment data be incorrectly uploaded to patient records.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- Sent an Urgent Medical Device Correction Safety Advisory Notice to customers on 1/11/11. Letters will be delivered by Siemens Customer Service Group or sent by certified mail. Letter identified the affected product and stated there is a potential risk due to a problem identified by the manufacturer. Customers are asked to refer to the attached Safety Notice from Elekta IMPAC Software. A Siemens Service Engineer will contact customers site to arrange for a patient matching criteria audit to determine if you have a problem. Customers should include the Safety Advisory Notice including the Safety Notice from Elekta IMPAC Software in their product related documentation until further notice. If customers have any questions, they should contact their responsible SIEMENS Service Engineer or Application Specialist.
- Quantity in commerce
- 10 units
- Distribution
- International Only Distribution -- Germany and Japan