Spectrum Central Venous Tray, Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter 7.0 Fr/...
Cook, Inc. began a recall of Spectrum Central Venous Tray, Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter 7.0 Fr/ 15cm, C-UTLMY-701J-ABRM-HC-FST-A-RD, G44126, Sterile... on . The reason FDA records is "Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1145-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Spectrum Central Venous Tray, Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter 7.0 Fr/ 15cm, C-UTLMY-701J-ABRM-HC-FST-A-RD, G44126, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
- Product code
- FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
- Reason for recall
- Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
- Quantity in commerce
- 51,473 total kits/trays
- Distribution
- Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
- Product codes and lots
- Part number: C-UTLMY-701J-ABRM-HC-FST-A-RD, Global part number G44126 Lot number: 2241830, 2241946, 2241947, 2244225, 2244835, 2244836, 2244837, 2244839, 2245830, 2253342, 2255605, 2255607, 2255738, 2255739, 2266266, 2267178, 2267969, 2268001, 2275493, 2276215, 2276880, 2276913, 2282406, 2282407, 2282422, 2282423, 2301416, 2301420, 2302127, 2304966, 2305905, 2307487, 2309428, 2313825, 2314047, 2315396, 2315399, 2318757, 2318842, 2321308, 2322431, 2322484, 2325091, 2326801, 2328402, 2328536, 2332350, 2335175, 2336043 and 2282407X