Spectrum Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter, 7.0 Fr/ 15...
Cook, Inc. began a recall of Spectrum Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter, 7.0 Fr/ 15 cm Power Injectable... on . The reason FDA records is "Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1191-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Spectrum Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter, 7.0 Fr/ 15 cm Power Injectable, C-UTLMY-701J-ABRM-HC-IHI-FST-A-RD, G49803, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
- Product code
- FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
- Reason for recall
- Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
- Quantity in commerce
- 51,473 total trays
- Distribution
- Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
- Product codes and lots
- Part number: C-UTLMY-701J-ABRM-HC-IHI-FST-A-RD, Global part number G49803. Lot numbers: 2240346; 2241834; 2245786; 2248322; 2249284; 2251999; 2253344; 2255176; 2255741; 2259342; 2268075; 2269800; 2276914; 2277908; 2281660; 2282559; 2285525; 2286943; 2291838; 2293485; 2294042; 2297529; 2299454; 2303178; 2306599; 2309430; 2311594; 2315346; 2316748; 2317453; 2318707; 2318718; 2318843; 2321268; 2321271; 2321282; 2321283; 2321299; 2325081; 2325083; 2325094; 2329811; 2331460; 2335177.
FIRST ARTICLE / WEEKLY