Spectrum Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter, 7.0 Fr/ 20...
Cook, Inc. began a recall of Spectrum Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter, 7.0 Fr/ 20 cm Power Injectable... on . The reason FDA records is "Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1193-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Spectrum Central Venous Tray Minocycline/Rifampin Impregnated Triple Lumen Polyurethane Catheter, 7.0 Fr/ 20 cm Power Injectable, C-UTLMY-701J-RSC-ABRM-HC-IHI-FST-RD, G49805, Sterile, Cook Medical, Cook Incorporated, Bloomington, IN.
- Product code
- FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
- Reason for recall
- Cook, Inc. is recalling Single, Double, Triple, and Five-Lumen Central Venous Catheter Trays AND Single and Double Lumen PICC Peripherally Inserted Central Venous Catheter Trays. These kits and trays contain 0.9% Sodium Chloride Injection USP 5 mL fill in 6 mL flush syringes that have been recalled by Excelsior Medical. During a routine test for syringe sterility, a possible integrity failure wa
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Cook, Inc. sent URGENT DEVICE RECALLS letters to its customers dated 10/4/2010. The letters asked customers to examine their inventory and notify Cook for product return. The letter included a response form.
- Quantity in commerce
- 51,473 total trays
- Distribution
- Worldwide distribution: USA, Chile, Colombia, Panama, Barbados, Canada, Guatemala, and Saudi Arabia.
- Product codes and lots
- Part number: C-UTLMY-701J-RSC-ABRM-HC-IHI-FST-RD, Global part number G49805. Lot numbers: 2289360; 2289361; 2289372; 2290112; 2294047; 2295470; 2295472; 2296081; 2296082; 2296096; 2297812; 2297813; 2297819; 2299200; 2300032; 2300736; 2301421; 2308256; 2308260; 2308355; 2311600; 2313815; 2317364; 2318712; 2318746; 2324271; 2326794; 2328390; 2328392; 2328397; 2328544; 232979; 2330540; 2330572; 2335187.
FIRST ARTICLE / WEEKLY