Z-1863-2011

Cardiac Marker Control Level 1, List Number 136601-06F12-02; Cardiac Marker Control Level 2, List number...

Abbott Medical / initiated 2010-12-17 / Terminated / root cause: design

Abbott Point Of Care Inc. began a recall of Cardiac Marker Control Level 1, List Number 136601-06F12-02; Cardiac Marker Control Level 2, List number 136602/06F13-02; and Cardiac Marker Control Level 3... on . The reason FDA records is "Presence of CO2 in the head space of the vial can result in the generation of high results in Cardiac Marker Control Level 1, 2, and 3 which is used along with the Cardiac Marker Control Calibration Verification Set". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1863-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Cardiac Marker Control Level 1, List Number 136601-06F12-02; Cardiac Marker Control Level 2, List number 136602/06F13-02; and Cardiac Marker Control Level 3, List number 136603/06F14-02. Used to verify the integrity of newly received i-STAT cTnI cartridges.
Product code
JJX Single (Specified) Analyte Controls (Assayed And Unassayed)
Reason for recall
Presence of CO2 in the head space of the vial can result in the generation of high results in Cardiac Marker Control Level 1, 2, and 3 which is used along with the Cardiac Marker Control Calibration Verification Set.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Recall Notice letters with business reply cards were sent Fed ex on December 17, 2010 to all accounts. International notifications were sent on December 21, 2010. The letter identified the affected product and lot numbers along with a background explanation of the issue. The firm recommends that customers remove the cap from the vial; squeeze the vial 10 times, prior to thawing; replace cap on the vial; and then follow the remaining steps for use as outlines in section 14-12 of the i-STAT 1 System Manual. Customers are to complete and return the Business Reply card to acknowledge receipt of the letter. If the affected product had been further distributed then a copy of the letter should be provided to those customers as well. If additional information is needed, customers should contact Abbott Point of Care Technical Support at 800-366-8020, Option 1 or their Abbott Point of Care representative.
Quantity in commerce
2360 units total
Distribution
Worldwide Distribution
Product codes and lots
Cardiac Marker Control Level 1 - Lot number M102081; Cardiac Marker Control Level 2 - Lot number M102082; and  Cardiac Marker Control Level 3 - Lot number M102083.